5 DUT Board Performance Considerations Correctly Terminating Signal Lines
122 System Reference, January 2001
Termination Checklist
50 Ohm Environment If you are operating in a standard 50 environment, no
extra action is needed as far as component layout on the
DUT board is concerned. You define the termination in the
Level setup.
Active Load If you terminate the line with the help of the active load,
no additional action is needed as far as component layout
on the DUT board is concerned. You define the active load
in the Level setup.
Voltage Clamp If you use the clamp load (useful for high-speed CMOS
devices, where 50 Ohm loading cannot be used), there is
no additional action needed as far as component layout on
the DUT Board is concerned. You setup the voltage clamp
in the Level setup.
The concept and the use of the active load and the voltage
clamp are discussed in the Standard Test Function Refer-
ence (Part No. E7050-91013).
Series Resistor If you use the resistive divider as a termination method,
you will need space on your DUT board for the series
resistor. For best results, use surface-mounted resistors
due to their low lead inductance.
For information on specifying the series resistor, refer to
the Command Reference manual (Part No. E7050-91013).