High Current Power Supply (HCDPS) 6 Device PowerSupply
System Reference, January 2001
179
disconnect is a hardware functionality which is only
available with the High Current DPS.
If this command is executed for the General Purpose
DPS or High Voltage DPS, it will have no effect.
The PSOP? query returns the last programmed value.
Syntax:
PSOP <code>,{(pinlist)}
PSOP? {(pinlist)}
returns
PSOP <code>,{(pinlist)}
Parameters:
<code>
0 Overvoltage protection (disconnect) disabled.
1 Overvoltage disconnect at U>Uset+0.2V.
2 Overvoltage disconnect at U>Uset+0.4V.
3 Overvoltage disconnect at U>Uset+0.6V.
4 Overvoltage disconnect at U>Uset+0.8V.
2. HSCM and HSCM?
Description:
The HSCM (High Current Clamp Mode) command is
similar to the PSCM command which sets the clamp
mode for the General Purpose and High Voltage Power
Supply channels.
The HSCM command has additional parameters for
determining the delay time for switching off the power
supply.
The HSCM? query returns the last programmed value.