Test System Configuration 4 Test Head Filling and DUT Board Considerations
System Reference, January 2001
73
Figure36 Group reservation of the digital dominant type 1024 pin SOC series
tester
m
The lightly shaded (yellow) groups in Figure 36 are the
only ones that can contain analog modules. Optionally,
you can have digital boards or analog modules in these
three groups. NOTE that Analog Module’s specifications
can only be guaranteed if these groups contain analog
modules only.
The more heavily shaded (dark grey) groups can only
contain digital boards.
There is a filling algorithm for filling the test head with
digital boards and analog modules:
• Digital Filling:
Group 1 ---> Group 2 ---> Group 3 ---> Group 4
---> Group 5 ---> (Group 7) ---> (Group 6) ---> (Group 8)
• Analog Filling:
(Group 8) ---> (Group 6) ---> (Group 7)
The bracket indicate that groups 6, 7 and 8 can optionally
be used for digital or analog testing. For this reason you
can find those groups under both list items.